Evident LogoOlympus Logo
Zdroje informací
Application Notes
Back to Resources

Vanta XRF Analyzer for Lead-Free Drinking Water


The US federal Safe Drinking Water Act (SDWA) was amended in 2011 to reduce the acceptable level of lead (Pb) to 0.25% in pipes, plumbing fittings and fixtures, solder, and flux that come into contact with potable drinking water. Effective in January 2014, the amended provisions require all suppliers involved in manufacturing, material handling, and installing brass plumbing components to comply with the Safe Drinking Water Act. X-ray fluorescence (XRF) is used to determine highly specific material chemistry, allowing rapid and accurate identification of metal grades for compliance with the Safe Water Drinking Act.

The Olympus Vanta XRF analyzer confirms alloy chemistry, including lead content and various other toxic elements, and provides a grade ID in seconds. The Vanta analyzer is used for simple sorting, incoming material inspection, and end-product verification.

Vanta XRF Analyzer for Lead-Free Drinking Water

Using XRF for EPA Lead Requirements

According to the Safe Water Drinking Act, all pipes, fittings, and fixtures manufactured and used in the US to carry potable water must be “lead free.” Section 1417 of the Safe Drinking Water Act establishes the definition for lead free as a weighted average of 0.25% lead calculated across the wetted surfaces of a pipe, pipe fitting, plumbing fitting, and fixture and 0.2% lead for solder and flux. Under the methodology established by this law, XRF is well suited to calculate average lead contact along the surfaces of pipes and plumbing fittings/fixtures.

Ideally, each individual component should be tested for lead. Testing for the presence of lead prevents material mix-ups and determines the levels of lead present in components. Studies have shown that the presence of chlorine in tap water accelerates the leeching of lead from plumbing components. It has also been shown that diffusion of lead into drinking water is correlated with lead concentration in piping and plumbing components.

The Vanta handheld XRF analyzer offers several advantages:

  • Rugged (IP55/54 rated and MIL-STD 810G drop tested)
  • Fast (readings in as little as 3 seconds)
  • Accurate (precision within 5% of measured value)
  • Quantify trace levels of lead (at and below 0.25%) and other toxic elements at trace levels
  • Confident grade identification for Envirobrass (SeBiLoys) and 125+ common copper grades
  • Identify alloy grades and verify material chemistry during incoming inspection, manufacturing, and final QC
Vanta XRF analyzer

Using the Vanta VCR model, we analyzed several brass and copper alloy certified reference materials (CRMs) for various heavy metals and regulated elements. The Vanta XRF analyzer’s excellent ability to detect low levels of lead in various EnviroBrasses for compliance with the 0.25% lead requirement is shown below. Quantification of lead was achieved at over 10 times below the action level. In addition, we analyzed an older, noncompliant brass sample for comparison.

Vanta XRF analyzer

Vanta (VCR) Reading
Alloy SeBiLoy Match? Pb (%) Pb +/- EPA Pass/ Fail (<0.25%)
EnviroBrass 2-1 YES 0.0343 0.0018 PASS
EnviroBrass 2-2 YES 0.0379 0.0021 PASS
EnviroBrass 2-3 YES 0.0221 0.0016 PASS
Older Plumbing Brass NO 0.2731 0.0015 FAIL

Additionally, the Vanta analyzer can effectively measure several other elements of concern for drinking water safety, including arsenic (As) and cadmium (Cd). Below are results from testing several brass CRMs for other toxic elements.

Vanta XRF analyzer

Vanta (VCR) Reading
Sample As Assay As (%) As +/-
As-1 0.025 0.0258 0.0013
As-2 0.029 0.0308 0.0012
As-3 0.0331 0.0371 0.0017
As-4 0.04 0.0494 0.0012
As-5 0.081 0.0909 0.0022
As-6 0.087 0.0959 0.0023
As-7 0.1 0.1057 0.0030
As-8 0.0137 0.0127 0.0014
As-9 0.057 0.0458 0.0021
As-10 0.041 0.0350 0.0017
As-11 0.014 0.0104 0.0013
As-12 0.03 0.0319 0.0012
As-13 0.046 0.0496 0.0017

Vanta XRF analyzer

Vanta (VCR) Reading
Sample Cd Assay Cd (%) Cd +/-
Cd-1 0.02 0.0259 0.0015
Cd-2 0.013 0.0197 0.0015
Cd-3 0.032 0.0372 0.0016
Cd-4 0.038 0.0424 0.0016
Cd-5 0.033 0.0392 0.0016

Olympus Vanta handheld XRF analyzers perform reliable positive material identification (PMI) and QA/QC of brass plumbing components to help users comply with the Safe Drinking Water Act. Vanta analyzers provide highly specific material chemistry to rapidly and accurately identify alloy grades and verify compliance to compositional requirements in seconds.

Vanta XRF analyzer

Olympus IMS

Produkty použité pro tuto aplikaci
Ruční XRF analyzátory Vanta pro analýzu půd a sedimentů poskytují rychlá a přesná data při vyhodnocování sanačních procesů. Analyzátory Vanta jsou jednoduché na zacházení a odolné i pro použití v náročných terénních podmínkách.
Ruční XRF analyzátory Vanta patří mezi nejodolnější přístroje, které Olympus vyvinul. Tato odolná, výkonná a intuitivní zařízení poskytují přesnou prvkovou analýzu a identifikaci slitin všem zákazníkům, kteří vyžadují v terénu výsledky laboratorní kvality.
Sorry, this page is not available in your country
Let us know what you're looking for by filling out the form below.
Sorry, this page is not available in your country