25 dicembre, 2014 Free replacement of Bräce-köhler compensator Thank you very much for your continuous business with us. Regarding our product of Bräce-köhler compensator ‘U-CBR1 / U-CBR2’, it was identified that the orientation setting of Gamma axis (Gamma axis : direction of slow optical vibration) was different from the indication of the products when shipping. We would therefore like to replace the applicable products free of charge. We sincerely apologize to our customers, our clients, and concerned parties for the great inconvenience caused by this issue. We will enhance our quality control system and make every effort to prevent any recurrence, and we would like to ask you for your understanding and cooperation. Olympus Corporation [Products in scope]Bräce-köhler compensator
[Problem]When measuring retardation, the direction of Gamma axis for a specimen is different from the direction indicated on the instruction manual by 90 degree. As a result of this, the checked result in determining the Gamma axis for the specimen will be different from the correct result by 90 degree. This problem does not affect the operation for the purpose of retardation measurement or enhancement of contrast.
[To identify the applicable products]
Find the serial number on the sticker placed on the inner part of the part indicating the product name.
[Request to our customers]
Our customers using the nonconforming item are kindly requested to contact the dealer where you purchased the item as detailed in [Contact information regarding this issue] below.
1. Product name (U-CBR1 or U-CBR2)
2. Serial number
3. Year, month of purchase (it is acceptable even if it is unclear) Please contact the dealer where you purchased the applicable product. [Contact information regarding this issue]Other Countries Please contact an agency of Olympus microscope around here.
* Applicable products include conforming items.
|
You are being redirected to our local site.