Did you know that you can use a Vanta XRF analyzer to accurately measure the thickness of a coating? If this is something you do regularly, setting up a coating model can help save you time. Creating the model is easy.

Select the right hand drop-down ‘gear’ symbol and select the ‘coating settings’ icon. When you do, you’ll see that silver (Ag) on copper (Cu) is the default. Select the ‘+’ icon to add a new model.

In the coating layer menu, write the name of the new model in the ‘name’ box. In this case, we’re calling it AgonCu5um. Below, there’s a drop-down menu that contains a list of elements from the alloy method beam 1 element suite. Select the appropriate element. In this case, we’re selecting silver on copper.

If your sample has multiple layers, select the ‘+’ icon to add another layer (you can create up to 3 layers). To remove a layer, simply double touch the trash icon.

It’s important note that when you’re using multiple layers, the analyzer counts the layer nearest the analyzer as the highest number layer.

Vanta measuring coating layers

Now we’re ready to conduct an optional one point calibration to help tune the analyzer to your sample. Press the ‘pencil’ icon. The default factor is set to 1, and the units are set to microns. Enter the target thickness for the standard of interest, and then select the ‘target’ icon to calibrate the instrument. The default time for the calibration is 10 seconds. Selecting a longer time will give you better precision. Select the ‘play’ icon to take the test.

Taking the test will recalculate the factor that has been set for the instrument and output the calculated thickness.

Without moving the sample, navigate to the live test screen and take a 10-second test. Notice that the coating name you selected appears at the top of the screen. The reported thickness will also show.

And that’s all there is to it!

Related Content

Through Thick and Thin: Using XRF to Measure Coating Thickness

Vanta Coating Mode Infographic

How to Add a Coating Template to Your Vanta Analyzer

Vanta Coating Mode Brochure


Get In Touch
학술 지원

Karen은 미국 매사추세츠주 월섬에 근거지를 둔 학술 고문입니다. 그녀는 2014년부터 Olympus에서 근무했으며, 신제품 개발을 지원하기 위해 Olympus X선 형광과 X선 회절 제품군을 지원하고 있습니다. XRF 및 XRD 장비를 유지하고 개선하는 데 전념하고 있으며, 고객과의 업무, 연구 개발 및 서비스를 담당하고 있습니다. Karen은 매사추세츠주 우스터에 있는 우스터 폴리테크닉 연구소에서 화공학 학사 학위를 취득했습니다. 


Karen has a Chemical Engineering degree from Worcester Polytechnic Institute in Worcester, Massachusetts. 







2월 23, 2018
죄송합니다. 이 페이지는 해당 국가에서 사용할 수 없습니다.
InSight 블로그 가입하기
죄송합니다. 이 페이지는 해당 국가에서 사용할 수 없습니다.
아래 양식을 작성하여 원하는 내용을 알려주십시오.
죄송합니다. 이 페이지는 해당 국가에서 사용할 수 없습니다.