Evident LogoOlympus Logo
Resources
Application Notes
Back to Resources

Meeting Strict Standards for Sulfur Content in Fuel with Vanta XRF Analyzers


Introduction

International and domestic regulations regarding sulfur content in petroleum products and fuels have become increasingly stringent in recent years. Sulfur needs to be removed from crude oil during the refinement process since it can impact vehicles’ emission control systems, contributing to air pollution. X-ray fluorescence (XRF) analysis can be used to measure the level of sulfur in fuel. Olympus’ Vanta XRF analyzer is well suited to meet the strict testing requirements established by standards, such as ASTM D-4294. The Vanta handheld analyzer can provide fast and accurate assessment of sulfur content in seconds and with virtually no sample preparation in even the most challenging environments.

Meeting Strict Standards for Sulfur Content in Fuel with Vanta XRF Analyzers

Using Vanta Analyzers for Low Sulfur Analysis

ASTM D-4294 establishes testing and performance guidelines when analyzing sulfur content using XRF in various petroleum products. The Vanta XRF analyzer offers several advantages for users who are testing to the D-4294 standard:

  • Rugged (IP55/54 rated and MIL-STD 810G drop tested)
  • Axon Technology electronics ensure accurate and repeatable performance
  • Virtually no sample prep —just collect your sample in a container and analyze it
  • The portable Vanta Work Station provides closed-beam safety and consistent sample presentation
  • Measure for other elements, such as wear metals (Fe, Cr, Ni, etc.) and additives (P, Ca, K, etc.)
 

To demonstrate the Vanta analyzer’s performance capabilities, we tested several certified oil samples for low sulfur content (<150 ppm) using a VCA (Vanta C series model with silver anode). Samples were tested inside the Vanta portable Work Station for 120 seconds each. Results shown are the averages of 5 repeated tests:

 
Meeting Strict Standards for Sulfur Content in Fuel with Vanta XRF Analyzers

Standard S Assay S S +/-
Mineral Oil 0 0 9
SMO-10-100 10 16 9
SMO-25-100 25 31,2 9
SMO-50-100 50 42,4 9
SMO-100-100 100 128,6 9
SMO-150-100 150 154 9

Conclusion

The Olympus Vanta handheld XRF can effectively and accurately measure sulfur content in accordance with ASTM D-4294. The accuracy and precision provided by the Vanta analyzer helps ensure that compliance with fuel regulations is maintained.

Olympus IMS

ProductsUsedApplications
Ręczne spektrometry XRF Vanta są jednymi z najbardziej odpornych urządzeń wyprodukowanych kiedykolwiek przez firmę Olympus. Dzięki wzmocnionej konstrukcji, dużej wydajności i intuicyjnemu oprogramowaniu, użytkownik otrzymuje szybką i dokładną analizę elementarną oraz identyfikację stopów, o jakości laboratoryjnej w warunkach terenowych.
Sorry, this page is not available in your country
Let us know what you're looking for by filling out the form below.
Sorry, this page is not available in your country