Evident LogoOlympus Logo

Решения НК
Specifications

Specifications are listed for all models unless stated otherwise

General and Enclosure

Size 335 mm × 221 mm × 151 mm (13.2 in. × 8.7 in. × 5.9 in.)
Weight 5.7 kg (with 1 battery) for model 16:64PR
5.9 kg (with 1 battery) for models 16:128PR, 32:128PR, and 64:128PR
Onboard Storage 1 TB internal SSD storage. Maximum 25 GB individual file size. Storage extensible using external USB drive
Storage Devices SDHC and SDXC cards, or most standard USB storage devices
Wireless Wi-Fi® 6E and Bluetooth® 5.3 onboard
PAUT Connectors 1 PA connector, 2x UT channels (2 P/R connectors each)
Available Configurations and Groups OmniScan X4A – 64 :128PR : 8 groups
OmniScan X4A – 32 :128PR : 8 groups
OmniScan X4A – 16 :128PR: 8 groups
OmniScan X4B – 16 :64PR: 2 groups (PA, UT, or TFM) or a specific configuration of 2 PA groups + 1 TOFD
Drop Testing Drop-tested according to MIL-STD-810G
Ingress Protection Rating IP65 Certified (completely protected against dust and water jets from all directions (6.3mm nozzle))
Display 269 mm (10.6 in.) TFT LCD with resistive touch screen, 1280 × 768 pixels
Battery 2 hot-swappable Lithium-ion batteries included, 87 Wh each
Battery Life Minimum of 5 hours (using ISO18563 standard setup parameters)
Operating Temperature −10 to 45 °C (14 °F to 113 °F)
Storage Temperature −20 °C to 60 °C (−4 °F to 140 °F) (with battery inside)
−20 °C to 70 °C (−4 °F to 158 °F) (with no battery inside)

Inputs and Outputs

Ports 2x USB 3.1, 1x video out (HDMI), 1x SDHC memory card slot, 1x communication (Ethernet)
Encoders 2-axis encoder line (quadrature or clock/direction)
Digital Inputs and Outputs 4 digital inputs (TTL), 4 digital outputs (TTL) including one reserved for acquisition on/off
External DC Supply 18 VDC, circular 2.5 mm diameter connector, center-positive

General Ultrasonic Specifications

Effective Digitizing Frequency Up to 100 MHz; User-adjustable compression factor
Maximum Pulse Repetition Rate (PRF) Up to 20 kHz; Actual maximum PRF is often limited by physics
(the time of flight required to obtain the ultrasonic signals)
Maximum Number of A-Scan Data Points Up to 16,384
A-Scan Bit Depth 16 bits
Rectification RF, full wave, half wave+, half wave−
Video Filtering Smoothing (adjusted to the probe frequency range)
Filtering PA channel: selection of band-pass filters, high-pass filters, and averaging
UT channel: selection of low-pass filters, band-pass filters, high-pass filters, and averaging
Time-Corrected Gain (TCG) PA: 40 dB per step of min. 0.1 dB
UT: 100 dB per step of min. 0.1 dB
Maximum slope of 40 dB/10 ns
Supported Beam Types (Groups) Selection of phased array ultrasonics (PAUT) beam types (linear, sectorial, and compound), single ultrasonic beam (UT), time-of-flight-diffraction ultrasonics (TOFD), total focusing method (TFM), phase coherence imaging (PCI), full matrix capture (FMC), and plane wave imaging (PWI)
Raw Full Matrix Capture (FMC) Data Collection Available only with model 64:128PR when using the NDT Device API or the OmniScan Black Box onboard application used with external acquisition software

Acoustic Specifications

 

 

PA Channel UT Channels
Certification Calibration Certification ISO 18563-1:2022 EN22232:2020

Pulse Width Adjustable from 30 ns to 1000 ns;
resolution of 5 ns (half period of bipolar pulse or duration of negative pulse)
Adjustable from 30 ns to 1,000 ns; resolution of 2.5 ns

Pulse Shape Bipolar negative-positive square pulse Negative square pulse
Receiver Gain Range 0 dB to 80 dB; maximum input signal 900 mVp-p (full-screen height). 0 dB to 120 dB; maximum input signal 30 Vp-p (full-screen height)

System Bandwidth 0.2 MHz to 26.5 MHz 0.25 MHz to 28.5 MHz

PAUT Specifications

OmniScan X4 Variant 64:128PR 32:128PR 16:128PR 16:64PR
Maximum Pulsing Aperture 64 32 16 16
Maximum Number of Receiving Channels 128 128 128 64
Number of Focal Laws 1024 maximum total (512 maximum per group)

FMC / TFM / PCI / PWI Specifications

OmniScan X4 Variant 64:128PR 32:128PR 16:128PR 16:64PR
Maximum Extended Aperture (FMC) 128 64 32 32
Maximum Aperture (PWI) 64 32 16 16
Maximum TFM or PCI Groups 4 4 4 2
Live TFM Envelope Yes – Actual Hilbert Transform
Image Resolution Up to 1024 × 1024 (1M points) for each TFM or PCI wave set
FMC Supported Wave Sets (TFM or PCI Modes) Pulse-Echo: L-L, T-T, and TT-TT
Self-Tandem: TT-T, TT-TTT, LL-L, LT-T, TL-T, TT-L, and TL-L
PWI Supported Wave Sets (TFM or PCI Modes) Pulse-Echo: L-L and T-T