2024년 6월 21일 Important Announcement for the SZ-FLR ESD Fluorescent Ring Illuminator for Stereo Zoom Microscopes Users Thank you very much for your continuous business with us. It has been discovered that the SZ-FLR ESD fluorescent ring illuminator for stereo zoom microscopes may cause electric shock due to leakage current if not grounded properly, although this is a very rare occurrence. We have been providing inspection and free repair services for this product since 2006. However, since 17 years have passed since we discontinued sales of this product and the maintenance period has also expired, we are sorry to inform you that we will no longer be providing this service.
To ensure safety, customers using the affected products are requested to dispose them.
We deeply apologize for any inconvenience this may cause to our customers, suppliers, and other stakeholders. We will make further efforts to prevent this from happening in the future and will take all possible measures for quality control. Affected ProductsFluorescent ring illuminator for stereo zoom microscopes (Electro-Static Discharge) The following products are not affected:
Cause and Description of FaultDue to some defects in the manufacturing process, leakage current may occur between the power supply unit and the lamphouse in very rare cases. If the product is used without grounding, there is a risk of electric shock. In case the product is properly grounded, the ground-fault breaker installed on the premises may trip, but there is no risk of electric shock. Request to CustomersCustomers using the affected products are requested to dispose of them safely. Appearance of SZ-FLR ESD, Location of Nameplate, Information in Nameplate
Appearance Bottom View(Name plate)
|
You are being redirected to our local site.